Vickers Hardness Number Calculator

By: Calculator Grid

Vickers Hardness Number Calculator

Convert a Vickers indentation load and mean diagonal into HV, surface-area hardness, and an empirical tensile-strength estimate.

Load 50 N Mean diagonal 0.5 mm Hardness 37.82 HV Estimated tensile strength 126.06 MPa
Workbook ready for the current example.

Indentation observations

Use a positive applied force and the arithmetic mean of the two measured indentation diagonals.

Accepted format: positive decimal with a dot; optional thousands commas.

ISO 6507-1 covers indentation diagonals from 0.020 mm to 1.400 mm.

Fixed test geometry: a 136° Vickers diamond pyramid. Standard gravity is 9.80665 m/s².

Live results

Results update as soon as both observations form a valid finite test state.

Vickers hardness number (HV) 37.82 HV · kgf/mm²
Surface area hardness (H) 370.87 MPa
Estimated tensile strength (σᵤ) 126.06 MPa
Indentation surface area 0.134817 mm²
Applied load 5.09858 kgf
Vickers hardness is 37.82 HV; estimated tensile strength is 126.06 MPa.

Calculation details

Each row is generated from the same canonical values used by the results panel and Excel workbook.

Quantity Calculation Current value
Applied load F ÷ g 5.09858 kgf
Indentation surface area d² ÷ [2 × sin(68°)] 0.134817 mm²
Vickers hardness number Load ÷ indentation surface area 37.82 HV
Surface area hardness HV × 9.80665 370.87 MPa
Estimated tensile strength HV ÷ 0.3 126.06 MPa
The tensile-strength figure is an empirical rule-of-thumb for metallic crystalline materials, not a substitute for a tensile test or a material-specific conversion standard.

How to use the Vickers hardness number calculator

What this calculator does

This calculator turns two observations from a Vickers indentation test – the applied Force (F) and the Mean diagonal length (d) – into four connected quantities. The main result is the Vickers hardness number (HV), expressed numerically in kilogram-force per square millimetre. It also reports Surface area hardness (H) in MPa, the calculated Indentation surface area, the converted Applied load in kgf, and an Estimated tensile strength (σᵤ). The hardness calculation is a geometric identity for the standard 136° Vickers indenter. The tensile-strength result is only an empirical screening estimate; it does not determine alloy grade, heat treatment, fracture behaviour, or certified tensile properties.

When to use it

Use the calculator when checking a laboratory worksheet after measuring a Vickers impression, comparing repeated indentations made at different loads, converting a force recorded in newtons into the conventional HV form, or preparing a quick engineering estimate before a fuller material-characterisation review. For conformity work, follow the applicable test method and equipment-verification requirements rather than relying on calculation alone. The current ISO 6507-1 Vickers test method specifies the metallic-material procedure and states a diagonal-length scope of 0.020 mm to 1.400 mm.

How to calculate

  1. The calculator opens with a complete demonstration: 50 N and 0.5 mm. Its results and a validated example Excel workbook are immediately available.
  2. Replace Force (F) with the load applied by the tester. Choose N or kgf; changing the unit converts the current value rather than relabelling it.
  3. Replace Mean diagonal length (d) with the arithmetic mean of the two measured indentation diagonals. Choose mm or µm; the value is converted automatically when the unit changes.
  4. Read Vickers hardness number (HV) first, then use the supporting results and the Calculation details table to audit the conversion path.
  5. Select Download Excel to create a fresh .xlsx workbook from the current validated values. Reset clears the demonstration data and results; Download Excel then remains disabled until both required observations are complete and valid again.

Input guide

Force (F) is required and accepts a positive decimal number with a dot as the decimal separator; correctly grouped thousands commas are also accepted. Select N for newtons or kgf for kilogram-force. A realistic example is 50 N. Increasing force while the measured diagonal stays unchanged increases HV in direct proportion. Do not enter a force unit inside the number field, use scientific notation, or type a decimal comma such as 1,5; ambiguous or unsupported formats are rejected visibly.

Force unit is required and controls how the entered force is interpreted. The conversion uses standard gravity, 9.80665 N per kgf. For example, 50 N becomes approximately 5.09858 kgf. Switching units preserves the physical load and therefore does not change the calculated hardness apart from display-level rounding.

Mean diagonal length (d) is required and must be positive. Enter the mean of both diagonals, not one diagonal selected because it looks cleaner. A realistic example is 0.5 mm. Because diagonal length is squared in the denominator, a smaller indentation gives a sharply higher hardness result, while a larger indentation gives a lower result. Optical resolution, surface preparation, edge proximity, and distorted impressions can dominate the uncertainty, especially for very small indents.

Diagonal unit is required and accepts mm or µm. One millimetre equals 1,000 micrometres, so the startup value 0.5 mm converts to 500 µm. Changing this selector preserves the same physical length. The UK National Physical Laboratory's overview of Vickers hardness testing describes the 136° diamond indenter and the measurement of the two residual diagonals.

Output guide

Vickers hardness number (HV) is the primary result. It equals the applied load in kgf divided by the sloping surface area of the indentation in mm². A higher value means greater resistance to the standardized plastic indentation under the stated conditions. Zero is not a valid result because both required inputs must be positive. HV is an indentation-hardness number and should not be treated as a universal pressure measurement or a guaranteed material identity.

Surface area hardness (H) converts the HV number to MPa by multiplying by standard gravity. It is driven by the same force and diagonal values and is an exact unit conversion of the calculated HV value. Indentation surface area is the geometric sloping area represented by the 136° pyramid; it rises with the square of the mean diagonal. Applied load is the current load converted to kgf, which makes the conventional HV equation auditable.

Estimated tensile strength (σᵤ) uses the rule of thumb HV ÷ 0.3 to report MPa. A higher hardness generally yields a higher estimate, but the relationship varies with alloy, microstructure, work hardening, and heat treatment. Treat it as an estimate, not an exact conversion. The four header pills repeat the current load, diagonal, HV, and tensile estimate for quick scanning. The Calculation details table shows each intermediate quantity, its equation, and its current value; the Excel workbook reproduces those same canonical numbers.

Worked example

With the startup values, F = 50 N and d = 0.5 mm. First convert the load: 50 ÷ 9.80665 = 5.09858 kgf. The indentation surface area is 0.5² ÷ [2 × sin(68°)] = 0.134817 mm². Dividing the load by that area gives 37.8186, displayed as 37.82 HV. Multiplying by 9.80665 gives 370.87 MPa surface area hardness. Dividing HV by 0.3 gives the displayed empirical tensile-strength estimate of 126.06 MPa. These numbers match the first-open results, table, and workbook checkpoints.

Learn more

Hardness results depend on more than arithmetic. Indenter geometry, force application, optical measurement, test-block calibration, and uncertainty all matter. NIST provides information on hardness standard reference materials used to check microhardness instruments, while ASTM's collection of indentation hardness testing standards identifies the relevant Vickers and microindentation methods.

Formula, sensitivity, and interpretation

HV = 2F sin(136° ÷ 2) ÷ (g d²) = 1.8544 × Fkgf ÷ d²

The squared diagonal is the most sensitive term. A 1% increase in force produces approximately a 1% increase in HV when diagonal length is held constant. A 1% increase in diagonal length produces approximately a 2% decrease in HV for small changes. That sensitivity is why image calibration, focus, corner selection, and averaging both diagonals are critical.

A single test state provides scalar quantities and unit conversions rather than a meaningful multi-point trend. For repeated tests, export each result and analyse the set with specimen identifiers, load, both individual diagonals, mean, HV, average, range, and standard deviation under the procedure required by your laboratory or specification.

A calculated HV value can be numerically correct while the physical test is invalid. Reject or investigate impressions affected by edges, neighbouring indents, poor surface preparation, cracking, excessive asymmetry, uncertain focus, or an unverified tester.